NIBP
  • Method Oscillometry
  • Modes Manual / Auto / Continuous
  • Auto Measure time Adjustable
  • Measurement range 10-270mmHg
ECG
  • Lead Mode 3-Iead Or 5-lead
  • Lead Selection I, II, III, avR, avL, avF, v
  • Heart rate range Adult: 15-300bpm, pediatric / Neonatal 15-300bpm
  • ECG waveform 2 channals
  • Accuracy ±1 bpm or±1 %, Whichever is greater
  • S-T segment detection: Measurement range 0.2mv~2.0mv
SPO2
Measurement range 0-100%
Accuracy 70-100%, ±2% 0-69%, unspecified
Pulse Rate
Range 20-254pm
Accuracy ±3bpm
Respiration
Method RA-LL impedance
Range Adult: 0-120rpm;pediatric/neonatal: 0-15rpm
Apnea alarm yes
Temperature
Range 0-50 °C
Accuracy ±0.1 °C
Channel Dual – channel
Specification
  • Alarm Yes, audible and visual alarm, alarm events review
  • High resolution 12.1” or 8.4” color TFT display
  • Lightweight, compact and portable
  • ECG, SPO2, NIBP, RESP, 2-TEMP, PR
  • Built-in rechargeable lithium battery
  • ECG waveforms of 7-leads display on the same screen
  • 72-hours graphic and tabular trends of all parameters
  • 72 alarm events of all parameters recall
  • 32 seconds full-disclosure waveform review
  • 500 NIBP measurement data can be stored and recall
  • Data and waveforms Manufacturer and Supplier color can be adjustable
  • Arithmetic analysis and S-T segment analysis
  • Suitable for adult, pediatric and neonatal patient
Optional
Nellcore SPO2, ETCO2, IBP, Printer & Networking
MP 1200 | Multi Parameter Monitor
Multipara Patient monitor Manufacturers, multi parameter patient monitor suppliers.
SPO2
Pulse Oximetry
Range 0% to 100%
Resolution 1 %
Accuracy 70% to 100% range: f 2% 0% to 69%
range: undefined
Method Dual wavelength LED
Configurations
Standard ECG, NIBP, SP02, HR,
Temperature * 2 and Respiration
Features
  • 12.1 ” color TFT LCD screen with multi-channel waveform display
  • Rechargeable high-energy built-in Li battery
  • Easy to learn, easy to use – simplified menus and dedicated Manufacturer and Supplier function keys
  • Easy connecting with central monitoring system
  • 96 hours data storage +Suitable for adult, pediatrics, neonates
ECG
Number of Leads 3- or 5-Lead
Lead View User-selectable; I, 11, Ill, aVR, aVL, aVF,
V (5-Lead); I, II, or Ill (3-Lead)
Gain Selection x114, x112, XaIn,d x2
Frequency Response: Diagnostic: 0.05 to 130 Hz
Monitor: 0.5 to 40 Hz
Surgery: 1 – 20Hz
Electrosurgery Protection: Yes
Defibrillator Protection: Yes
Pacer DetectionlRejection: Yes
Temperature
Channels 2
Range, Accuracy 71.6″F to 122°F (28°C to 50°C) : f0.IoC
Display Resolution f O . l OC
Respiration Rate
Rate 7 to 120 bpm (ECG)
Resolution 1 breath / min
Accuracy +2 breath / min
Trends
Trend 1 Hour resolution I s or 5 s
Display 72 Hour resolution I min, 5mins, 10 mins
tabular, graphical
NIBP (Non-invasive Blood Pressure)
Technique Oscillometric during inflation
Range Adult: 40 to 270 mmHg
Pediatric 40 to 200 mmHg
Neonate 40 to 135 mmHg
Measurement Cycle < 40 sec. typical
Automatic Measurement Cycles (Selectable)
1, 2, 3, 5, 10, 15, 30 min; 1, 2, 4, 6 hr
STAT Mode 5 minutes of continuous readings
Max. Allowable Cuff Adult: 300 mmHg
Pressure Pediatric: 240 mmHg
Neonate I50 mmHg
Resolution 1 mmHg
Transducer Accuracy f 3 mmHg
Heart (Pulse) Rate
Source User selectable: Smart, ECG, Pleth, NlBP
Range NIBP: 40 to 240 bpm
ECG 15 to 300 bpm(Adult)
15 to 350 bpm(Neonate)
SP02 20 to 300 bpm
Accuracy f I bpm or f I % (ECG) whichever is greater
f 3bpm(SP02, NIBP)
Power Requirements
Voltage 100-250 V AC; 50160 Hz
Power Consumption 70 Watts, typical
Battery lead battery
Battery Life more than 100 minutes
Recharge Time less than 12 hours
Optional
Masimo SPO2, ETCO2, IBP & Networking, Printer, Touch Screen
MP 1210 | Multi Parameter Monitor
Multipara Patient Monitor, Multi Para Moniter in India
NIBP
  • Method Oscillometry
  • Modes Manual / Auto / Continuous
  • Auto Measure time Adjustable
  • Measurement range 10-270mmHg
SPO2
Measurement range 0-100%
Accuracy 70-100%, ±2% 0-69%, unspecified
ECG
  • Lead Mode 3-Iead Or 5-lead
  • Lead Selection I, II, III, avR, avL, avF, v
  • Heart rate range Adult: 15-300bpm, pediatric / Neonatal 15-300bpm
  • ECG waveform 2 channals
  • Accuracy ±1 bpm or±1 %, Whichever is greater
  • S-T segment detection: Measurement range 0.2mv~2.0mv
Pulse Rate
Range 20-254pm
Accuracy ±3bpm
Respiration
Method RA-LL impedance
Range Adult: 0-120rpm;pediatric/neonatal: 0-15rpm
Apnea alarm yes
Temperature
Range 0-50 °C
Accuracy ±0.1 °C
Channel Dual – channel
Specification
  • Alarm Yes, audible and visual alarm, alarm events review
  • High resolution 12.1” or 8.4” color TFT display
  • Lightweight, compact and portable
  • ECG, SPO2, NIBP, RESP, 2-TEMP, PR
  • Built-in rechargeable lithium battery
  • ECG waveforms of 7-leads display on the same screen
  • 72-hours graphic and tabular trends of all parameters
  • 72 alarm events of  Manufacturer and Supplier all parameters recall
  • 32 seconds full-disclosure waveform review
  • 500 NIBP measurement data can be stored and recall
  • Data and waveforms color can be adjustable
  • Arithmetic analysis and S-T segment analysis
  • Suitable for adult, pediatric and neonatal patient
Optional
Nellcore SPO2, ETCO2, IBP, Printer & Networking

Leave a Reply

Your email address will not be published. Required fields are marked *

Related Posts